Power
Chem or Petrochemical
Pulp and Paper
Semiconductor
Cation exchange breakthrough,for high acid samples,increase the effective run time on cation exchangers,drum boiler water
. Sensitivity to ppb range
. Advanced embedded MCU technology,SMT and integrative design
. Panel mount,aluminum enclosure
---anti-interference
. 192×64 lattice LCD display
. User interface language:English
. Simultancous display of multtiparameters
. Automatic or manual temperature compensation
. Two analog outputs and two alarm relay outputs
. CAN bus communication
. Large datalog of previous measurements and trend curve trackable
. Universal power supply for worldwide operation(100~240)V
. Low reagent consumption
. Fast response,accurate measurement
---ease-of-use with minimal maintenance
TRANSMITTER |
|
Measurement Range |
(0.00 to 9.36)pNa, 0.01μg/L to 23.0g/L, (-1.000 to +1.000)mV |
Resolution (min) |
0.01pNa, 0.01ug/L, 0.1mV |
Accuracy |
±0.01pNa, 1character, ±0.5mV |
Stability |
±0.01pNa/24h, ±1character/24h, ±0.5mV/24h |
Analog Output |
(0~10)mA or (0~20)mA or (4~20)mA,two isolated outputs |
Output Accuracy |
±1%F.S |
Max.load Resistance |
≤800W |
Power Supply |
(100~240)VAC, (45~65)Hz |
Power Consumption |
≤8W |
Insulation Strength |
Power to ground 1500V, no flying arc in 1 min |
Alarm Output |
Two alarm relays. Each relay can be configured as: high concentration alarm, low concentration alarm, high temperature alarm and low temperature alarm. One relay can be simultaneously selected by two alarms. If anyone of the two alarms occurs, it will be activated. |
Communication |
CAN Bus (option) |
Data Logger |
Record measurement values, history events and calibration results |
INSTRUMENT |
|
Measurement Range |
(1.00-8.00)pNa, 0.23ug/L-2300mg/L |
Resolution (min) |
0.01pNa or 0.01ug/L |
Accuracy |
±0.02pNa or ±0.5ug/L ( the larger one), whichever is greater (between ±10oC of calibration temperature) |
Response Time (T90) |
90%within 2 min |
Temperature Sensor |
Pt1000 |
Temperature Range |
(0.0 to 99.9)oC |
Temperature Accuracy |
±0.5 oC |
Temperature Compensation |
(0.1 to 60.0)oC (auto or manual) |
Calibration Method |
Sample calibration and standard solution calibration |
Overall Dimension (mm) |
450(L) × 690(W) × 190(D) |
Cutout Dimensions (mm) |
410 × 645 |
Weight |
19kg |
Sample Condition |
Temperature: (5 to 50)oC Pressure: (50-300)KPa Flow: (6-40)L/h Solid contents in sample: ≤5mm (no gel and grease) |
Ambient Temperature |
(5 to 45)oC |
Ambient Humidity |
≤90%RH (no condensation) |
Storage Temperature |
(-25 to 55)oC (above 0oC for electrode) |
Warranty |
18 months(from date of acceptance) |
Analyzer |
1piece |
Measuring Electrode |
1piece |
Reference electrode |
1piece |
Standard solution |
2bottle |
Reactive solution |
1bottle |
Displsable injector |
1piece |
Filling solution for reference electrode |
1bottle |
Connector |
1piece |
Screw driver |
1piece |
F46 pipe |
1piece |
Key |
1piece |